Q: Which filling solution do I use with my Princeton Applied Research reference electrode?
普林斯頓的參比電極需要使用哪一種填充液?
A: Listed below are the references electrodes sold by Princeton Applied Research and their proper filling solutions. For your convenience, part numbers are provided for all of the filling solutions available. For certain electrodes, filling solutions are not offered. This is because the solution is easily made by the user (and costing less than the shipping charges if we shipped it), custom made by the user, or should be freshly made by the user.
Reference Electrode | Filling Solution | Filling Solution Chemistry |
K0077 Saturated Calomel | - - - - - - - * | Saturated KCl |
219995 Ag/AgCl | 1601-0167-0 | KCl/AgCl (Sat’d in both), 0.5 oz |
K0265 Ag/AgCl | SL0070 | 3M NaCl/Saturated AgCl, 1.0 oz |
G0093 Hg/HgSO4 | - - - - - - - * | Saturated K2SO4 |
K0103 Non-Aqueous | G0155 | 0.1 M AgNO3 in acetonitrile, 125 mL |
Ag/AgCl** | RDE0022 | AgCl/KCl (Sat’d in both), 125 mL |
G0159 Reference E'trode Jacket*** | RDE0022 | AgCl/KCl (Sat’d in both), 125 mL |